Description
Application
SEA-13P1 is used in nuclear physics centers for research in various fields of fundamental and applied physics, in radiochemical laboratories of enterprises for monitoring technological processes, in laboratories of external dosimetry services of enterprises for monitoring environmental objects, in radiological laboratories of the State Sanitary and Epidemiological Service, in veterinary and agricultural services for monitoring the quality of water of centralized and non-centralized water supply, animal products and soils.
Design and operation
SEA-13P1 is a single unit that includes an evacuated chamber with a silicon ion-implanted semiconductor detector, a spectrometric amplifier, a spectrometric ADC, and a power source for the spectrometer and detector.
The spectrometer also includes a vacuum pump, a personal computer with printer, and SpectraLineADA spectrometry software.
Features and functional capability
- Samples of aerosol emissions collected on AFA-RSP-20 aerosol filters can be measured without special radiochemical treatment.
- Samples to be measured must be prepared by electrolytic deposition
- Samples to be measured are placed on one of ten shelves in the measuring chamber, located at distances from 5 to 50 mm from the detector in 5 mm increments.
- Maximum sample diameter is 70 mm